EasyManua.ls Logo

Keysight PXA N9030A - Page 607

Keysight PXA N9030A
1138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9 Strip Chart Measurement
Remote Command Results
Command N Results Returned
INITiate:SCHart n/a n/a
CONFigure? n/a name of current measurement: " SCHart "
CONFigure:SCHart n/a n/a (selects Strip Chart measurement in Meas Preset state)
INITiate:SCHart n/a n/a (selects Strip Chart measurement without affecting settings)
FETCh:SCHart [n]?
MEASure:SCHart [n]?
READ:SCHart [n]?
Not specified
or n=1
Return the following comma-separated scalar results:
1. Current measurement result of Meter 1
2. Current measurement result of Meter 2
3. Current measurement result of Meter 3
Note : Return NaN (9.91E37) if the Meter is not available.
n=2 Return the following comma-separated scalar results:
1. Max measurement result of Meter 1
2. Max measurement result of Meter 2
3. Max measurement result of Meter 3
Note : Return NaN (9.91E37) if the Meter is not available.
n=3 This query returns Trace 1 data as a list of x,y pairs. The y-values are in the
current Y-Axis Unit of the analyzer. The x-axis values are the values of
the trace, in the x-axis scale units of the trace (Hz for frequency domain
traces, seconds for time domain traces).
n=4 Returns Trace 2 data as a series of x,y pairs
n=5 Returns Trace 3 data as a series of x,y pairs
Key Path
Meas
Initial S/W Revision A.07.00
EMI Receiver Mode Reference 607