Loading...
9 Strip Chart Measurement
Meas Setup
25 µs
Sets the De-emphasis time constant to 25µs.
Key Path
Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis
Example SCH:DEM:FM:DEEM US25
Readback
25 s
Initial S/W Revision A.07.00
50 µs
Sets the De-emphasis time constant to 50 µs.
Key Path
Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis
Example SCH:DEM:FM:DEEM US50
Readback
50 s
Initial S/W Revision A.07.00
75 µs
Sets the De-emphasis time constant to 75 µs.
Key Path
Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis
Example SCH:DEM:FM:DEEM US75
Readback
75 s
Initial S/W Revision A.07.00
750 µs
Sets the De-emphasis time constant to 750 µsec.
Key Path
Meas Setup, Analog Demod Tune&Listen, FM, De-emphasis
Example SCH:DEM:FM:DEEM US750
Readback
750 s
Initial S/W Revision A.07.00
EMI Receiver Mode Reference 657