EasyManua.ls Logo

Keysight U1281A - Calibration Process; Test Considerations

Keysight U1281A
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 Calibration Procedures
26 Keysight U1281A/U1282A Service Guide
Calibration Process
1 Prior to performing the verification tests, see the “Test Considerations” on
page 26.
2 Perform the verification tests to characterize the multimeter; see “Performance
Verification Tests” on page 27.
3 Unsecure the multimeter for calibration; see “Calibration Security” on page 37.
4 Prior to performing the adjustments, see the Adjustment considerations” on
page 45.
5 Perform the adjustment procedure; see Adjustment procedure” on page 53.
6 Secure the multimeter against unauthorized calibration; see “Exiting the
adjustment mode” on page 54. Ensure that the multimeter has quit the
adjustment mode and is turned off.
7 Record the new security code and calibration.
Test Considerations
For optimum performance, all procedures should comply with the following
recommendations:
The performance verification test or adjustment should be performed under
laboratory conditions where the ambient temperature can be controlled.
The multimeter should be put under the laboratory environment for at least 1
hour.
Ensure that the calibration ambient temperature is stable and is between 18 °C
and 28 °C. Ideally the calibration should be performed at 23 °C ± 1 °C.
Ensure that the ambient relative humidity is less than 80%.
Allow a warm-up period of 5 minutes.
Use a shielded twisted pair of PTFE-insulated cables to reduce settling and
noise errors. Keep the input cables as short as possible. Long test leads can
also act as antennas which may pick up AC signals.
Connect the input cable shields to earth ground.

Table of Contents

Other manuals for Keysight U1281A

Related product manuals