1 Calibration Procedures
46 Keysight U1281A/U1282A Service Guide
Valid adjustment input values
Adjustment can be accomplished using the following input values.
Table 1-19 Adjustment input values
Test function Step Reference value Valid reference input
AC V
6.00 V
3.00 V (55 Hz) 0.9 to 1.1 x Reference value
3.00 V (1 kHz) 0.9 to 1.1 x Reference value
0.30 V (55 Hz) Reference Value
3.00 V (55 Hz) 0.9 to 1.1 x Reference value
3.00 V (10 kHz) 0.9 to 1.1 x Reference value
60.00 V
3.00 V (55 Hz) Reference value
30.00 V (55 Hz) 0.9 to 1.1 x Reference value
30.00 V (1 kHz) 0.9 to 1.1 x Reference value
30.00 V (10 kHz) 0.9 to 1.1 x Reference value
600.00 V
30.00 V (55 Hz) Reference value
300.00 V (55 Hz) 0.9 to 1.1 x Reference value
300.00 V (1 kHz) 0.9 to 1.1 x Reference value
300.00 V (10 kHz) 0.9 to 1.1 x Reference value
1000.00 V
30.00 V (55 Hz) Reference value
300.00 V (55 Hz) 0.9 to 1.1 x Reference value
300.00 V (1 kHz) 0.9 to 1.1 x Reference value
300.00 V (10 kHz) 0.9 to 1.1 x Reference value
Vsense
[a]
Hi.10 V 10.00 V (55 Hz) High Sense
Lo.30 V 30.00 V (55 Hz) Low Sense