1 Calibration Procedures
36 Keysight U1281A/U1282A Service Guide
Table 1-17 Temperature performance verification test
Test function Range
5520A/5522A
output
Error from nominal 1 year
U1281A U1282A
Temperature
While the rotary is in the position, press
.
-200 °C to 1372 °C 0 °C
± (Probe
accuracy +
1 °C)
± (Probe
accuracy +
1 °C)
Esc
Shift
NOTE
Set both Fluke 5520A and multimeter to internal reference.
To perform the measurement, connect one end of the K-type thermocouple
(with miniature TC connector on both ends) to the 5520A TC output, and the
other end to the multimeter via a TC-to-banana adapter. Allow at least
1 hour for the multimeter to stabilize.
Table 1-18 Frequency counter performance verification test
Test function Range 33522B output
Error from nominal 1 year
U1281A U1282A
Frequency counter
While the rotary switch is in the position, press
.
9.9999 MHz 1 MHz, 200 mV N/A ±0.0005 MHz
99.999 MHz 20 MHz, 330 mV N/A ±0.0054 MHz
MHz
Esc
Shift