Calibration Procedures 1
Keysight U1281A/U1282A Service Guide 57
Calibration Error Codes
The following errors indicate failures that may occur during a calibration.
Table 1-20 Calibration error codes
Error code Description
200/E200 Calibration error: calibration mode is secured
002/E002 Calibration error: secure code invalid
003/E003 Calibration error: serial number code invalid
004/E004 Calibration error: calibration aborted
005/E005 Calibration error: value out of range
006/E006 Calibration error: signal measurement out of range
007/E007 Calibration error: frequency out of range
008/E008 EEPROM write failure