Calibration Procedures 1
Keysight U1281A/U1282A Service Guide 29
Table 1-4 AC V (V) performance verification tests
Test function Range
5520A/5522A
output
Error from nominal 1 year
U1281A U1282A
AC V (V)
Turn the rotary switch to the position and press
.
6 V
6 V, 20 Hz ±0.0660 V ±0.0660 V
6 V, 45 Hz ±0.0205 V ±0.0205 V
6 V, 1 kHz ±0.0205 V ±0.0205 V
6 V, 10 kHz ±0.0445 V ±0.0445 V
6 V, 20 kHz ±0.0960 V ±0.0960 V
6 V, 100 kHz N/A ±0.3160 V
60 V
30 V, 20 Hz ±0.360 V ±0.360 V
60 V, 45 Hz ±0.205 V ±0.205 V
60 V, 1 kHz ±0.205 V ±0.205 V
60 V, 10 kHz ±0.445 V ±0.445 V
60 V, 20 kHz ±0.960 V ±0.960 V
60 V, 100 kHz N/A ±3.160 V
600 V
30 V, 20 Hz ±0.90 V ±0.90 V
600 V, 45 Hz ±2.05 V ±2.05 V
600 V, 1 kHz ±2.05 V ±2.05 V
600 V, 10 kHz ±4.45 V ±4.45 V
300 V, 20 kHz ±5.10 V ±5.10 V
320 V, 30 kHz ±12.70 V ±12.70 V
1000 V
1000 V, 45 Hz ±5.5 V ±5.5 V
1000 V, 1 kHz ±5.5 V ±5.5 V
1000 V, 10 kHz ±9.5 V ±9.5 V
300 V, 20 kHz ±10.5 V ±10.5 V