EasyManua.ls Logo

Keysight U1281A - Table 1-4 AC V (V) Performance Verification Tests

Keysight U1281A
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Procedures 1
Keysight U1281A/U1282A Service Guide 29
Table 1-4 AC V (V) performance verification tests
Test function Range
5520A/5522A
output
Error from nominal 1 year
U1281A U1282A
AC V (V)
Turn the rotary switch to the position and press
.
6 V
6 V, 20 Hz ±0.0660 V ±0.0660 V
6 V, 45 Hz ±0.0205 V ±0.0205 V
6 V, 1 kHz ±0.0205 V ±0.0205 V
6 V, 10 kHz ±0.0445 V ±0.0445 V
6 V, 20 kHz ±0.0960 V ±0.0960 V
6 V, 100 kHz N/A ±0.3160 V
60 V
30 V, 20 Hz ±0.360 V ±0.360 V
60 V, 45 Hz ±0.205 V ±0.205 V
60 V, 1 kHz ±0.205 V ±0.205 V
60 V, 10 kHz ±0.445 V ±0.445 V
60 V, 20 kHz ±0.960 V ±0.960 V
60 V, 100 kHz N/A ±3.160 V
600 V
30 V, 20 Hz ±0.90 V ±0.90 V
600 V, 45 Hz ±2.05 V ±2.05 V
600 V, 1 kHz ±2.05 V ±2.05 V
600 V, 10 kHz ±4.45 V ±4.45 V
300 V, 20 kHz ±5.10 V ±5.10 V
320 V, 30 kHz ±12.70 V ±12.70 V
1000 V
1000 V, 45 Hz ±5.5 V ±5.5 V
1000 V, 1 kHz ±5.5 V ±5.5 V
1000 V, 10 kHz ±9.5 V ±9.5 V
300 V, 20 kHz ±10.5 V ±10.5 V