1 Calibration Procedures
46 Keysight U3606B Service Guide
4-wire resistance gain verification test
1 Connect the calibrator to the front panel (red) and LO (black) input
terminals and (red) and (black) output terminals (see Figure 1-5, “Test
setup for 4-wire resistance gain verification,” on page 31).
2 Press [] to select the resistance function.
3 Select each range in the order shown below. Provide the indicated input
resistance. Compare the measurement results to the appropriate test limits in
Table 1-8 (be certain to allow for appropriate source settling when using the
Fluke 5520A).
Table 1-7 2-wire resistance gain verification test
Input resistance Range Error from nominal 1-year
100 Ω 100 Ω
±0.058 Ω
[a]
[a] Specifications stated are for 2-wire resistance measurements using the Null math operation.
Without null, add an additional error of 0.2 Ω.
1 kΩ 1 kΩ
±0.00055 kΩ
[α]
10 kΩ 10 kΩ
±0.0055 kΩ
[α]
100 kΩ 100 kΩ ±0.055 kΩ
1 MΩ 1 MΩ ±0.00065 MΩ
10 MΩ 10 MΩ ±0.0255 MΩ
100 MΩ 100 MΩ ±2.005 MΩ
For the 1 Ω and 10 Ω range, please perform calibrator zeroing before testing the
resistance values of the device-under-test (DUT) to ensure that the 5520A is
maintained with a zero cal. every 12 hours ± 1
o
C.