Calibration Procedures 1
Keysight U3606B Service Guide 47
Frequency voltage gain verification test
1 Connect the calibrator to the front panel V (red) and LO (black) input terminals
(see Figure 1-7, “Test setup for frequency voltage gain verification,” on
page 32).
2 Press [Hz ms %] to select the frequency function.
3 Select each range in the order shown below. Provide the indicated input
frequency and voltage. Compare the measurement results and the appropriate
test limits in Table 1-9 (be certain to allow for appropriate source settling when
using the Fluke 5520A).
Frequency current gain verification test
1 Connect the calibrator to the front panel I (red) and LO (black) input terminals
(see Figure 1-8, “Test setup for frequency current gain verification,” on
page 32).
2 Press [Hz ms %] to select the frequency function.
Table 1-8 4-wire resistance gain verification test
Input resistance Range Error from nominal 1-year
1 Ω
[a]
[a] Calibrator ohms zero cal. every 12 hours ± 1
o
C
1 Ω ±0.0028 Ω
10 Ω
[a]
10 Ω ±0.012 Ω
100 Ω 100 Ω ±0.12 Ω
1000 Ω 1000 Ω ±1.2 Ω
Table 1-9 Frequency voltage gain verification test
Input frequency Input voltage Range Error from nominal 1-year
1 kHz 1 V 1.999 kHz ±0.00026 kHz