4.3. Perform averages
During measurement, it is possible to perform an automatic average. This automatic average is calculated on the basis of
one measurement each seconds. It is also possible to get an addition of several dataset of automatic average.
4.3.1. Perform an automatic average
Press "Moy/Avg".
Press OK or the red button on the right of the base to launch measurement.
Measurement duration appears at the lower right of the screen.
Press OK or the red button on the right of the base to display the results when required measurement duration is
reached.
Screen displays the following results in the left column:
Average
Minimum value
Maximum value
Deviation: standard deviation on all the measured values
Press up and down arrows to get to results of the different parameters.
Press “Save” to save the dataset.
Screen backs to the display of results with:
Left column: measurement results
Right column: addition of averages on several datasets of automatic averages
Results displayed in both columns are identical because based on a single dataset.
4.3.2. Perform an aggregate of automatic averages
Results of the aggregate of averages are based on several datasets of automatic average, it means that the DBM 610
should not be turn off and the “Esc” key should not be pressed.
Once a 1
st
dataset of automatic average has been realized and saved:
Press OK or the red button on the right of the base to go to measurement screen.
Press OK again or the red button on the right of the base to launch a new measurement.
Press OK again or the red button on the right of the base to display results when required measurement duration is
reached.
Press up and down arrows to get to results of the different parameters.
Screen displays the following results in the left column:
Average
Minimum value
Maximum value
Deviation: standard deviation on all the measured values
No result is displayed on the right column (Σ).
Press up and down arrows to get to results of the different parameters.
Press “Save” to save the dataset.
On measurement screen:
Press "Esc" to cancel and not include these results. It will be possible to launch a new measurement.
Press "Save" to save results.
Screen displays the following results in the left column:
Left column: measurement results
Right column (Σ): aggregate of averages on several datasets of automatic averages
Using DBM610 in airflow meter mode 13

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kimo DBM610 and is the answer not in the manual?

Kimo DBM610 Specifications

General IconGeneral
BrandKimo
ModelDBM610
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals