14.8. Guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration - Electromagnetic emissions
The device KINETEC
®
Breva is intended for use in the electromagnetic environment specified below. The customer or the user of the device
KINETEC
®
Breva should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
Radio frequency emissions CISPR 11 Group 1
The device KINETEC
®
Breva uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Radio frequency emissions - CISPR 11 Class B
The device KINETEC
®
Breva is suitable for use in all establishments including
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions - IEC 61000-3-2 Class A
Voltage fluctuations / Flicker emissions - IEC
61000-3-3
Complies
Guidance and manufacturer’s declaration - Electromagnetic immunity
The device KINETEC
®
Breva is intended for use in the electromagnetic environment specified below. The customer or the user of the device
KINETEC
®
Breva should ensure that it is used in such an environment.
Immunity test
IEC 60601
Test level
Compliance level Electromagnetic environment – Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient /
burst
IEC 61000-4-4
±2 kV
for power supply lines
±2 kV
for power supply lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±0,5 kV, ±1 kV
between lines
±0,5 kV, ±1 kV, ±2 kV
between line and earth
±0,5 kV, ±1 kV
between lines
±0.5 kV, ±1 kV, ±2 kV
between line and
earth
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage interruptions
IEC 61000-4-11
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
< 5% U
T
(>95% dip in U
T
)
for 5 seconds
When the interruption occurs, the device KINETEC
®
Breva
is reset. After turning on, push START to begin the session.
Voltage dips
IEC 61000-4-11
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
device KINETEC
®
Breva requires continued operation
during power supply interruptions, we recommend
powering the device KINETEC
®
Breva using an
uninterruptible power supply or a battery.
Power frequency (50/60
Hz) magnetic field - IEC
61000-4-8
3A/m 3A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.