Instructions for Use
Electrosurgical Unit Minicutter
Revision 08 21
Risk of malfunction and danger of injury from high current concentration!
After prolonged application of HF current at high power, the surface of the device can heat up significantly.
In
case of HF activation for more than 15 s, an acoustic warning signal is sounded by increasing the volume
of the activation sound. The activation sound always accompanying HF activation is an important safety
element. It can be regulated on the rear of the
unit, but cannot be deactivated.
Immediately stop activation of the foot switch pedal.
The current type bipolar coagulation is deactivated.
Risk of malfunction, danger of injury!
If the HF activation signal can be heard while there is no foot switch
connected, the device is faulty:
The device must not be operated. A technical check is required in any such case.
Risk of malfunction, danger of injury!
If the HF activation signal can be heard while the foot switch is connected but the pedal is not
pressed,
the
foot switch is defective:
This accessory must not be used. It must be replaced.