Instructions for Use
Electrosurgical Unit Minicutter
Revision 08 35
10 Currents, their Properties and Technical Data
Cutting 1 (Pure)
Clean cutting without eschar formation
Power setting range: Max. 80 W at 300
Max. periodic HF output voltage: max. 1,200 Vmax
Max. output current 510 mA at 300
Crest factor 2.1
Danger of injury from excessive voltage!
Active accessories to be used with this current must be able to withstand a combination of an HF voltage
of at least
1,200 V. It must be kept in mind that voltage is achieved even at settings below 10, and that the
active accessory must be selected accordingly.
Performance Diagram
HF power in W
Setting