EasyManua.ls Logo

KLS Martin Minicutter - Page 43

KLS Martin Minicutter
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use
Electrosurgical Unit Minicutter
Revision 08 43
Bipolar coagulation
Coagulation current with high voltage for surface coagulation (fulguration).
Power setting range: Max. 70 W at 50
Max. periodic HF output voltage: Max. 340 V
max
Max. output current 1.200 mA at 50
Crest factor 2.5
Danger of injury from excessive voltage!
Active accessories to be used with this current must be able to withstand an HF voltage of at least 340
V.
It
must be kept in mind that voltage is achieved even at settings below 10 and that the active accessory
must be selected accordingly.
Performance Diagram
HF power in W
Setting

Table of Contents

Related product manuals