Recommended separation distances between portable and mobile HF telecommunications equipment and the
Minicutter
Requirements for high-frequency wireless communication equipment
Frequency band
(MHz)
Test frequency (MHz) Modulation Compliance level (V/m)
380 – 390 385 Pulse
a
– 18 Hz 27
430 – 470 450
FM ± 5 kHz sweep or
pulse
a
– 18 Hz
28
704 – 787 710, 745, 780 Pulse
a
– 217 Hz 9
800 – 960 810, 870, 930 Pulse
a
– 18 Hz 28
1700 – 1990 1720, 1845, 1970 Pulse
a
– 217 Hz 28
2400 – 2570 2450 Pulse
a
– 217 Hz 28
5100 – 5800 5240, 5500, 5785 Pulse
a
– 217 Hz 9
Note
A minimum safety distance of 30 cm (12″) between portable HF communication
equipment emitting in the respective frequency band and the Gebrüder Martin Minicutter
should be observed. This includes e.g. cell phones and WIFI, RFID and Bluetooth devices.
Non-observance may result in reduction of the performance features of the device.
a