EasyManua.ls Logo

Knick Protos 34009(X) - Page 24

Knick Protos 34009(X)
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
Protos 3400(X)
SW 3400-009: Concentration Table
Select menu: Parameter setting/System control/Concentration table
Specifying a concentration solution for conductivity measurement
Concentration table
To specify the customer-specific solution, 5 concentration values A-E are
entered in a matrix together with 5 temperature values 1-5.
To do so, first enter the 5 temperature values, then enter the respective
conductivity values for each concentration A-E.
These solutions will then be available in addition to the permanently stored
standard solutions (select Table”).
Menu Display
Entering a concentration table
To enter values
Open parameter setting
System control
Select “Concentration table
Enter 5 temperature values
(right/left arrow keys to select
position, up/down arrow keys to edit
number, confirm by pressing enter.)
Abort
25.6 °C
0.020 mS/cm
Temperature 2
Temperature 3
Temperature 1
Concentration table (Administrator)
Enter values for concentrations A-E
for the respective temperatures. The
table values must be continuous.
Maxima/minima are not permitted.
Incorrect entries are marked with
.
The concentration table is selected as follows:
Parameter setting/Module COND/Cal preset values/Automatic/Table.
Temperature 4
Temperature 5
+005.0 °C
+010.0 °C
+015.0 °C
+020.0 °C
+000.0 °C
OK
Abort
OK
Return
25.6 °C
0.020 mS/cm
Point of measurement
Logbook
Factory setting
Release of options
Calculation Blocks
Concentration table
System control (Administrator)
25.6 °C
0.020 mS/cm
Concentration A:
3. Cond at +010.0 °C
4. Cond at +015.0 °C
2. Cond at +005.0 °C
5. Cond at +020.0 °C
1. Cond at +000.0 °C
Concentration table (Administrator)
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
05.00 % by wt

Related product manuals