25
The table is built up as 5 x 5 matrix:
Conc. A Conc. B Conc. C Conc. D Conc. E
Temp 1 A1 B1 C1 D1 E1
Temp 2 A2 B2 C2 D2 E2
Temp 3 A3 B3 C3 D3 E3
Temp 4 A4 B4 C4 D4 E4
Temp 5 A5 B5 C5 D5 E5
Conditions for the table:
The temperature values must be rising
(Temp 1 is the lowest, Temp 5 the highest temperature).
The concentration values must be rising
(Conc. A is the lowest, Conc. E the highest concentration).
The table values A1 ... E1, A2 ... E2, etc. must all be rising within the table or
all falling. Points of inflection are not allowed!
The analyzer automatically checks the table values.
Incorrect entries are marked.