1-4 2G0733 January 14, 2008
Overview
Automatic Image Quality Control
An internal densitometer is a key element in the Automatic Image Quality 
Control (AIQC) process. The densitometer enables the Imager to 
automatically adjust image processing parameters to produce the best image. 
The Imager adjusts these parameters each time it prints a calibration film.
A calibration film is printed when:
• The film tray is inserted in the Imager with film of a new lot number.
• A calibration film is requested from the local panel or web portal.
• A film tray is inserted into the Imager for which a current calibration 
is not stored.
Agency Compliance
See the KODAK DRYVIEW 5800 Laser Imager Safety Manual, 2G0734.
User Guide Conventions
The following special messages emphasize information or indicate potential 
risks to personnel or equipment.
NOTE: Notes provide additional information, such as expanded 
explanations, hints, or reminders.
IMPORTANT: Important notes highlight critical policy information that 
affects how you use this guide and this product.
Cautions point out procedures that you must follow precisely to 
avoid damage to the system or any of its components, loss of data, 
or corruption of files in software applications.
DANGER: Danger identifies procedures that you must follow 
precisely to avoid injury to yourself or others.
Laser warnings warn personnel that access to laser radiation is 
possible and all personnel must avoid direct exposure to the beam.