5-2 2G0733 January 14, 2008
Specifications
Other Location 
Considerations
Additional factors that influence where the Imager is located include 
proximity to patients, flammable materials, liquids, and other equipment.
This equipment is not contained in a sealed cabinet. Do not use this 
equipment in locations where it can come in contact with liquids, 
including body fluids.
IMPORTANT:
• This equipment is intended to connect to other medical devices. Only 
an authorized Service Provider of Carestream Health, Inc. Products or 
Customer Qualified Service Personnel may install this equipment.
• Do not locate the Imager within 2 m (6.56 ft.) of where a cell phone 
will be used, even if the cell phone is separated from the equipment 
by a wall.
• Do not locate the Imager within 4 m (13.12 ft.) of a microwave oven. 
Electromagnetic radiation from a microwave oven is only an issue if, 
after the oven door is closed and latched, the seal does not maintain 
an electromagnetic tight fit between the oven door and the oven main 
housing. Determining if the seal has an electromagnetic tight fit 
requires special detection equipment.
• Do not locate the Imager in the presence of flammable anesthetics, 
oxygen, or nitrous oxide. The Imager does not have a gas-sealed 
electronics enclosure and could ignite any flammable or explosive 
gases present in the Imager’s environment.
• Do not locate the Imager in close contact with MRI devices due to 
possible high magnetic fields near an MRI unit. The magnetic field in 
the area where the Imager is installed must be less than 50 Gauss.
• Do not locate the Imager closer than 1.83 m (6 ft.) from a patient bed 
or chair.
• Do not substitute or modify any part of the Imager without prior 
written approval of Carestream Health, Inc.