EasyManua.ls Logo

Kodak Dryview 5800 - Environmental Requirements; Temperature; Relative Humidity; Altitude

Kodak Dryview 5800
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
January 14, 2008 2G0733 5-3
Environmental Requirements
Temperature Operating: 15° to 33° C (59° to 91° F)
Storage: -40° to 60° C (-40° to 140° F)
Relative Humidity Operating: 20% to 80% RH, noncondensing
Storage: 10% to 90% RH, noncondensing
Altitude 30 m (100 ft.) below sea level to 3,048 m (10,000 ft.) above sea level
Surface Condition The surface must be a level (must be level within 1°) table top or counter
capable of supporting the weight of the Imager. For the Imager to operate
with no image degradation, surface vibration levels cannot exceed the
following:
Environmental Effects
Heat Dissipation: total heat dissipation for the Imager is 850
BTU/hour
Acoustical Noise:
Less than or equal to 75 dB-A at 1 m during a sound burst of 1
sec. or less
Less than or equal to 65 dB-A at 1 m during normal operation
Less than or equal to 50 dB-A at 1 m during idle mode
Laser Specifications
Wavelength - 810 nm
Power - 50 mW
Radiation class - 3B
Type - Diode
G
2
/Hz
Freq. Range Direction
1 x 10
-7
1-50 Hz Vertical or horizontal
6 x 10
-5
50-200 Hz Vertical or horizontal
5 x 10
-6
200-650 Hz Vertical or horizontal
1 x 10
-6
650-1000 Hz Vertical or horizontal