EasyManua.ls Logo

Kodak EKTAPRO EM 1012 - Imager Remote Record;Stop

Kodak EKTAPRO EM  1012
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HIGH GAIN
NORMAL
RECORD
STOP
MONITOR
HEADSET
Record / Stop Connector
1
3
2
ACCESSORIES
The RECORD pin and the STOP pin supply +5 Volts DC at
approximately 100 micro-amperes.
The GROUND pin can be connected to the RECORD pin OR the
STOP pin with a dry contact switch closure or an electronic
circuit capable of sinking 300 micro-amperes.
IMAGER REMOTE RECORD/STOP
The RECORD/STOP connector on the bottom of the Imager
allows you to start and/or stop a recording remotely. A mating
connector is supplied with the Imager or a replacement can be
ordered using our Part Number 03100075-301. The solder side
of the mating connector is shown below.
Wiring the RECORD/STOP connector.
The RECORD/STOP connector pin out is:
1 = RECORD
2 = GROUND
3 = STOP