4-12 A-61504 October 2007
Use the calibration target provided with your scanner.
1. Click Calibrate. The Calibrate dialog box will be displayed.
2. Select Image Chain or UDDS from the drop-down box.
If you select Image Chain, use the calibration target provided
with your scanner. Be sure to use a good, clean calibration target.
Place the calibration target in the input elevator face up and click
OK.
IMPORTANT:If the White Background Accessory is installed, it
must be replaced with the black background strips,
and the scanner must be restarted prior to
calibration.
If you select UDDS, place a 20 lb. or 75 g/m
2
bond sheet of paper
in the input elevator. Do not use the calibration target and click
OK.
3. Click Start. While calibration is being performed a status will be
provided in the Status box.
4. Click Done when the calibration is successful.