24
The display shows:
Phenacetin
detected: 135.3*
* Please note that due to the factory adjustment, a
slightly different value for the melting point may be
displayed.
Set the new value for phenacetin and then for
sulfanilamide and caffeine according to the previous
procedure.
Save all changes by pressing the print button.
The system menu will be displayed. To exit to the main
menu press the print button again.
Notes!
Please note that for the adjustment of the M5000
only melting points should be used, which were also
determined with the M5000.
In principle, when using different melting point meters
from different manufacturers, slight deviations of the
determined melting points can occur. This would
also be noticeable in the adjustment.
4. Calibration with reference substances