9. Subsequent retrieval of the last melting point
actual temperature
25.8 °C
9. Subsequent retrieval of the last melting point
If you missed the time for reading the melting point and
the device has returned to the starting display, the last
melting point can be displayed once more by pressing
the printing key.
melting point
121.9 °C
The display will appear for about 5 sec. After that the
device will return to the starting display. If the printing
key is activated once more within the 5 s, the print-out
will be created with the attached printer (if available).
This can be repeated as often as you want.
The term “auto” will not be shown on the display, since
no melting point analysis can be started in this case
inserting a capillary tube does not have any effect.
If an internal error value is determined as melting point,
the melting point analysis was prematurely aborted by
pressing the arrow keys (clear text display of the error
for 5 s).
The melting point could not be determined due to a
time-out (10 min.), i.e. the melting point could be
10 °C higher than the selected starting temperature
(clear text display of the error for 5 s).
abort,
no melting point