8
●Display: Liquidcrystaldisplay
Insulationresistancerange:(Max.1200counts)
VAC/DCrange:(Max.630counts)
Bargraph:Max.36points
DAR/PIvalue:Max.9.99
Time:Max.99:59
●Lowbatterywarning: Batterysymbol(in4levels)
●Overrangeindication: OLmark app ears on insulation
resistancerange.
Himarkappearsonvoltagerange.
●Auto-ranging:
Rangeshiftstoupperrange:1000counts
Rangeshiftstolowerrange:80counts
(Onlyontheinsulationresistancerange)
●Auto-power-off:
Theinstrumentwillbepoweredoffifthere
isnoswitchorbuttonoperationfor10min.
(This function does not work during a
measurement.)
●Altitude: 2000morless
●Temperature&humidityrange(guaranteedaccuracy):
23ºC±5ºC/Relativehumidity85%orless(nocondensation)
●Operatingtemperature&humidityrange:
0ºCto40ºC/Relativehumidity85%orless(nocondensation)
●Storagetemperature&humidityrange:
-20ºCto+60ºC/Relativehumidity75%orless(nocondensation)
●Overloadprotection:
Insulationresistancerange:AC1200V/10sec.
Voltagerange:AC720V/10sec.
●Withstandvoltage: AC5160V(50/60Hz)/5sec.
(Betweenelectricalcircuitandenclosure)
●Insulationresistance:1000MΩormore/DC1000V
(Betweenelectricalcircuitandenclosure)
●Dimension: 177(L)×226(W)×100(D)mm
●Weight:
KEW3125A:approx.1.9kg(batteryincluded),
KEW3025A:approx.1.7kg(batteryincluded)
●Powersource: DC12V: LR14(Alkaline battery size C)x
8pcs
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30