20
3. How to measure DAR/ PI
DAR and PI are automatically measured during a normal
continuousmeasurement of insulation resistances. Set the
Range switch to any desired range and measure the test
objectcontinuously.
-1minafterastartofcontinuousmeasurement:
LCDshowsDARvalue.
-10minafterastartofcontinuousmeasurement:
LCDshowsPIvalue.
WhenDAR/PIvaluesaredisplayedasno:
DAR and PI values are determined by Methods 1. and 2.
as described above, therefore, they are displayed as no
whenthemeasuredinsulationresistancesfallunderanyofthe
followingcases.
*1measuredvalueis0.0MΩ
*2measuredvalueisOL
*OLis displayed when the measured value exceeds
the upper limit of the measuring range at each insulation
resistancerange.
Range Upperlimit
250V 105.0MΩ
500V 1050MΩ
1000V 2.10GΩ
2500V 105.0GΩ
5000V 1200GΩ
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30