EasyManuals Logo

Labconco FreeZone Plus User Manual

Labconco FreeZone Plus
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
Appendix C: Freeze Dryer Specifications
Product Service 816-333-8811 or 1-800-522-7658
47
Environmental Conditions
Indoor use only.
Maximum altitude: 6562 feet (2000 meters).
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F
(31°C), decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient overvoltages according to Installation Categories II
(Overvoltage Categories per IEC 1010). Temporary voltage spikes on
the AC input line that may be as high as 1500V for 115V models and
2500V for 230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally
only non-conductive atmospheres are present). Occasionally,
however, a temporary conductivity caused by condensation must be
expected, in accordance with IEC 664.
Heat rejected from freeze dyer excluding the vacuum pump: Less
than 4320 BTU/HR.

Table of Contents

Other manuals for Labconco FreeZone Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco FreeZone Plus and is the answer not in the manual?

Labconco FreeZone Plus Specifications

General IconGeneral
BrandLabconco
ModelFreeZone Plus
CategoryCommercial Food Equipment
LanguageEnglish

Related product manuals