Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658
19
Original instructions
6. Lift-Away™ front panel provides easy access to electrical wiring, sash
weights, and lighting fixtures.
7. Energy efficient LED lighting is located behind a laminated safety glass
shield mounted to the top of the hood. The factory-wired instant start T8
lighting is serviceable from outside the hood cavity. Additionally, the
long lasting 50,000 hour direct wired LED T8 bulbs are more energy
efficient, utilizing approximately ½ the power of fluorescent bulbs. See
Appendix A for wattage.
8. Low mounted, factory-wired light and blower switches are ADA
compliant.
9. All Air Foils allow air to sweep the work surface for maximum
containment. Airfoils are offered with either Eco-Energy Saving or Flush
Foil. The Clean-Sweep™ openings create a constant barrier from
contaminants. In addition, should the operator inadvertently block the
airflow entering, the air continues to enter from under the air foil and
through the Clean-Sweep openings. See Appendix A for alternate PVC
Eco-Air Foil as PVC material resists corrosion from mineral acids.
10. Streamlined corner posts provide maximum visibility and the flexibility
to add services after installation.
11. All hoods are factory prepared for up to 8 service fixtures.
12. Duplex electrical receptacles are mounted on the right and left corner
posts as requested. Receptacles are factory-wired to hood single point
junction box.
13. Shipped fully assembled and eliminates the need for costly onsite
assembly.
14. Accessory Guardian™ Digital Airflow Monitor or Guardian Airflow
Monitor continuously monitors face velocity. An audio/visual alarm
alerts the user to low airflow conditions. The right corner post is factory
prepared to accommodate the Guardian Monitor (sold separately).
15. Optional Sash Models. Sashes are offered in smooth anti-racking cable,
chain, auto-return, or Intelli-Sash.
16. Frame of epoxy-coated steel and aluminum is durable and corrosion
resistant.
17. Exhaust Connection. The hood features 12" (12.75" OD pipe) exhaust
connections sized to allow for a minimum static pressure loss through the
hood structure while providing a good transport velocity through the
exhaust system. 4' Integral Blower models use 10" (10.75" OD).
18. Spillstopper™ Solid Epoxy Work Surface is dished to contain spills.
(Work surface is sold separately).
19. Optional Ceiling Enclosure Kits are available for a decorative facade
between the hood and the ceiling. Integral Blower models have a facade as
shown in Figure 4-1.

Table of Contents

Questions and Answers: