Product Service 1-800-522-7658
45
Original instructions
APPENDIX C
PROTECTOR PREMIER
SPECIFICATIONS
Environmental Conditions
Indoor use only.
Maximum altitude: 10,000 feet (3,048 meters).
Ambient temperature range: 41° to 104°F (5° to 40°C).
Maximum relative humidity: 80% for temperatures up to 88°F (31°C),
decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage fluctuations not to exceed ±10% of the nominal
voltage.
Transient over-voltages according to Installation Categories II (Over-
voltage Categories per IEC 1010). Temporary voltage spikes on the AC
input line that may be as high as 1500V for 115V models and 2500V for
230V models are allowed.
Used in an environment of Pollution degrees 2 (i.e., where normally only
non-conductive atmospheres are present). Occasionally, however, a
temporary conductivity caused by condensation must be expected, in
accordance with IEC 664.
Electrical Ratings
Volts AC
Phase
Cycle
AMP*
115
1
50/60
10
115
1
50
10
115
1
60
10
230
1
50/60
5
230
1
50
5
230
1
60
5
*Does not include current rating of receptacles.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Protector Premier 10040 Series and is the answer not in the manual?

Labconco Protector Premier 10040 Series Specifications

General IconGeneral
BrandLabconco
ModelProtector Premier 10040 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals