50
PROBLEM CAUSE CORRECTIVE ACTION
Defective hose to fill
switch
Check for splits.
Check for loose clamps.
Defective fill level
switch
Recalibrate switch. Contact Labconco Product
Service.
Replace switch.
Defective controller Check output from PCB to both valves and DI
pump.
Replace PCB. Contact Labconco Product
Service.
Holes in filter screen
too small
Contact Labconco Product Service.
Defective grommet
where hose connects
to fill level switch
Replace grommet.
Low Temp Alarm
Heater inoperative Check electrical connections.
Check voltage to heater.
Check resistance of heater. If open circuit,
replace heater.
Defective controller Check output from PCB to heater.
Replace PCB. Contact Labconco Product Service.
Defective
temperature probe
Check electrical connections.
Contact Labconco Product Service. Replace if
necessary.
Defective high
temperature cut out
switch
Check resistance of safety cut out switch. If open
circuit at room temperature, replace switch.

Table of Contents

Other manuals for Labconco SteamScrubber 44003 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco SteamScrubber 44003 Series and is the answer not in the manual?

Labconco SteamScrubber 44003 Series Specifications

General IconGeneral
BrandLabconco
ModelSteamScrubber 44003 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals