12. Press *IOK to select Scan Sample. Turbidity (F)
* Scan Sample
16:02:19 01/04/05
13. Observe the result. Turbidity (F)
0.54 NTU
* Scan Sample q
16:02:19 01/04/05
14. Press and then press *IOK to
select Calibrate.
Turbidity (F)
0.54 NTU
* Calibrate q
16:02:19 01/04/05
15. Use the or to change the
highlighted digits on the display to
match the concentration of the
turbidity standard. Press *IOK to
accept a digit and move to the
next digit.
Calibrate
00.54
q , Q , p
16:02:19 01/04/05
Calibrate
00.54
q , Q , p
16:02:19 01/04/05
Calibrate
00.54
q , Q , p
16:02:19 01/04/05
Calibrate
00.50
q , Q , p
16:02:19 01/04/05
ANALYSIS & CALIBRATION 27