CHAPTER 5
MAINTENANCE OF THE WASHER
NU_810-815_LX_ANG_1306
59
5. PREVENTIVE MAINTENANCE
GETINGE LANCER’S TECHNICAL ASSISTANCE SERVICE recommends to all the users at least
one annual preventive maintenance visit.
The preventive maintenance reduces the risks of stops of the washer due to breakdowns and
enables enlarge the life duration of the equipments.
Depending on the countries where the washer is used and the local norms, a higher frequency of
preventive maintenance visits can be necessary.
6. PREVENTATIVE MAINTENANCE SCHEDULE
FUNCTION
Daily
Biannual
1 Year
2 Year
5 Year
CLEAN FILTER SYSTEM IN CHAMBER.
X X X X X
CHECK CHEMICAL CONTAINERS FOR ANY LEAKAGE.
X X X X X
ENSURE CHEMICAL HOSES ARE NOT PINCHED.
X X X X X
CLEAN EXTERIOR PANELS.
X X X X X
INSPECT WATER SUPPLY HOSES FOR CRACKS, BULGES, AND LEAKS.
X X X X
ENSURE THE WATER HOSE SEALS AND FILTERS ARE CLEAN AND HAVE NO CRACKS.
X X X X
CHECK CHEMICAL SUPPLY HOSES FOR CRACKS, BULGES AND LEAKS.
X X X X
INSPECT INTERNAL TUBING ON CHEMICAL PUMPS FOR WEAR.
(I
F APPLICABLE)
X X X X
CHECK CHEMICAL LEVEL SENSORS FOR CORRECT OPERATION.
(I
F APPLICABLE)
X X X X
ENSURE ALL PANELS ARE PROPERLY SECURED.
X X X X
CHECK ALL INTERNAL HOSES FOR CRACKS, BULGES AND LEAKS.
X X X X
ENSURE ALL HOSE CLAMPS ARE PROPERLY TIGHTENED.
X X X X
CHECK ALL COLUMN SEALS FOR LEAKS AND CRACKS.
X X X X
CHECK SPRAY ARM SUPPORT SEALS FOR LEAKS AND CRACKS.
X X X X
CHECK SPRAY ARMS AND BUSHINGS FOR WEAR.
X X X X
CHECK DOOR SEAL AND GASKET FOR LEAKAGE.
X X X X
INSPECT THE DOOR SPRINGS, DOOR WHEELS AND THEIR SLING, HOOKS, DOOR SWITCH FOR
PROPER OPERATION
.
X X X X
CHECK RECYCLING AND EMPTYING PUMP SEALS FOR LEAKAGE AND QUIET OPERATION.
X X X
CHECK FOR LOSE ELECTRICAL CONNECTIONS AT COMPONENTS AND ELECTRONIC CARDS.
X X X
VERIFY THE CORRECT OPERATION OF ALL RELAYS AND THEIR ASSOCIATED COMPONENTS.
X X X
CHANGE THE HEATING RELAY.
X
(1) Daily maintenance must be handled by users staff.
(2) Others maintenances (bi-annual, 1 year, 2 years & 5 years) must be carry out by the
GETINGE LANCER’s technical assistance service.