19
TROUBLE CAUSE REMEDY
Water only dispensed, no
syrup. Or syrup only dispensed,
no water
1. Syrup BIB empty
2. Water or syrup shuto on mounting block
not fully open
3. Improper or inadequate water or syrup
supply
4. CO2 pressure to syrup pump too low
5. Stalled or inoperative BIB pump
6. Kinked line
7. CO2 regulator malfunction
8. Defective LFCV module
1. Replace syrup BIB as required
2. Open shuto completely
3. Remove valve from mounting block &
o p e n s h u t o s s l i g h t l y . C h e c k w a t e r & s y r u p
supply. If no supply, check unit for other
problems. Ensure BIB connection is
engaged
4. Check the CO
2
pressure to the pump to
ensure it is between 70-80 PSI (0.482 -
0.552 MPA)
5. Check CO
2
pressure and/or replace pump
6. Remove kink or replace line
7. Repair or replace CO
2
regulator as
required
8. Replace module
Syrup only dispensed. No
water, but CO
2
gas dispensed
with syrup
1. Improper water ow to dispenser
2. Carbonator pump motor has timed out
(display message on the LCD screen)
3. Liquid level probe not connected prop-
erly to PCB
4. Defective PCB assembly
5. Defective liquid level probe
6. Weak or defective carbonator pump
1. Check for water ow to dispenser
2. Reset by turning the unit OFF, then ON
by using the circuit breaker on the power
supply or momentarily unplugging unit
3. Check connections of liquid level probe to
PCB assembly
4. Replace PCB assembly
5. Replace liquid level probe
6. Replace pump
Excessive foaming 1. No ice in bin
2. Ice not touch cold plate
3. Incoming water or syrup temperature too
high
4. CO
2
pressure too high
5. Water ow rate too high
6. Nozzle and diuser not clean
7. Air in BIB lines
1. Fill bin with ice and allow cold plate to
re-stabilize
2. Increase auto agitation time
3. Correct prior to dispenser
4. Adjust CO
2
pressure downward, but not
less than 70 PSI (0.482 MPA)
5. Re-adjust and reset ratio
6. Remove and clean
7. Bleed air from BIB lines