Proline Kryomats
YAUE0007 / 2/26/2009 Maintenance 103
Master: Alarm message al/B
Command: Pump blocked
(Pump motor monitoring: Overload,
blockage).
Ì 7.12.5.
1.
The viscosity of the heat terans
fer liquid is too
high Æ change heat carrier liquid or raise the
setpoint temperature.
2. The pump is blocked. Æ Contact the LAUDA
Service Constant Temperature Equipment Ì 9.4.
Master: Alarm message OslDT
Command: Low level (pump)
(Pump motor monitoring: No load).
Ì 7.12.6.
1.
No liquid in the system. If this oc
curs, the level
monitoring has failed. Æ Check whether the float
in the expansion vessel is blocked by foreign
bodies. Otherwise, contact LAUDA Service
Constant Temperature Equipment Ì 9.4.
2.
With the option "open load"
the device draws air
out of the open load. Æ Move the return to the
load.
Master: Alarm message Dppmp 00
Command: Overpressure
(outflow pressure too high).
Pump level too high Æ Select a lower pump level
Ì 7.5.3.
Compressor overtemperature Ì 7.12.7
Three-phase current Ì 7.12.8

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda PROLINE Kryomats Series and is the answer not in the manual?

Lauda PROLINE Kryomats Series Specifications

General IconGeneral
BrandLauda
ModelPROLINE Kryomats Series
CategoryThermostat
LanguageEnglish

Related product manuals