56 Starting up YAUE0007 / 2/26/2009
The temperature measurement device
shows the true temperature value (with
glass thermometers consider the
correction where applicable!).
Change the display in the adjacent
window to the true value with cursor or
soft keys and accept with or End ,
or quit the window with without making
changes.
7.6.7 Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
Default
Open the device parameter menu via
the soft key Menu .
With the cursor keys continue to
Æ Settings Æ Calibration Æ
intern Pt100 Æ Default .
The window shown adjacent appears.
Confirm selection with .
Confirm the control dialog on the right
with or cancel with .
Return to the measurement window with
End or .
Hel
p
Menu Screen T
set
T
fix
6 7 8 9 0
1 2 3 4 5
Temperature value of the ref.
temp. measurement device:
20.15
Min: -50.0°C Max: 302.00
Hel
p
Menu End T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Hel
p
Menu End T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel