Proline Kryomats
YAUE0007 / 2/26/2009 Starting up 57
7.6.8 Entering the offset of the external temperature probe
If a deviation is found during the check using a calibrated reference thermometer, e.g. from the LAUDA
DigiCal Series, then the offset (the additive part of the characteristic) of the external measurement chain
can be adjusted with the following function. The reference thermometer must be dipped nearly by the
external temperature probe into the consumer bath according to the details on the calibration certificate.
Command
− Calibration
− Open the device parameter menu via the
soft key Menu .
− With the cursor keys continue to
Æ Settings Æ Calibration Æ
extern Pt100 Æ Calibration .
− The adjacent window appears.
− Confirm selection with .
− Continue as described in Ö 7.6.6 for the
internal temperature probe.
7.6.9 Restoring the works setting of the external temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
− Default
− Open the device parameter menu via the
soft key Menu .
− With the cursor keys continue to
Æ Settings Æ Calibration Æ
extern Pt100 Æ Default .
− The adjacent window appears.
− Confirm selection with .
− Continue as described in 7.6.7 for the
internal temperature probe.
Hel
Menu End T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Hel
Menu End T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default