YAUE0007 / 2/26/2009 Starting up 53
7.6.3 Setpoint offset operating mode
With this function it is possible to apply an offset value to the temperature provided by the external
temperature probe or a module and then to use it as the setpoint. The bath temperature can, for
example, be operated at -25 °C below the temperature of a reactor, which is being measured by the
external temperature probe.
Command
Offset source and Setpoint offset
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Æ Control Æ Setpoint offset Æ Offset
source .
Off indicates that the setpoint offset is
currently deactivated.
Select the setpoint source with or
and confirm with .
Interfaces (e.g. RS232) are only
displayed if a valid setpoint has already
been transmitted.
With the cursor keys continue to
Æ Setpoint offset .
The standard value is 0.00°C .
Open the left-hand window with .
Enter the desired temperature.
Accept the change with ,
quit the window with without making
changes.
Hel
p
Menu End T
set
T
fix
Off
extern Pt100
RS232
Offset source
Setpoint offset
Hel
p
Menu Screen T
set
T
fix
6 7 8 9 0
1 2 3 4 5
Input Setpoint offset:
0.00
Min: -500.00°C Max: 500.00°C
Hel
p
Menu End T
set
T
fix
0.00 °C
Offset source
Setpoint offset