EasyManuals Logo

Lauda RP 890 User Manual

Default Icon
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
Proline Low-temperature Thermostats
62 Starting up YACE0072 / 21.08.07
7.8.3 Defining temperature limits
With this function, it is possible to define a minimum and a maximum temperature in which the
thermostat controls. By reaching the temperature limits, a warning appears. In this way setpoint input
can be prevented which may damage the heat carrier medium or the apparatus. For example, if water is
used as the heat carrier liquid, +95°C would be practicable as the maximum temperature and +5°C as
the minimum temperature.
Master
rhg and rhK
3x to LMDls
to
O@p@ 3x
Call the start option
rhg (maximum temperature),
or with 4x
rhK (minimum temperature).
The maximum temperature can be changed here.
The minimum temperature can be changed here.
The maximum temperature (minimum temperature) is displayed
flashing.
or
Set the required limit temperature .
Wait 4 seconds or
Display flashes 4 s Æ new value is automatically accepted, or
Value is immediately accepted with the Enter key.
Command
Limits
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to Limits .
The minimum and maximum
temperatures are displayed.
T il (min) is currently active.
Select the limit to be changed with or
and confirm with .
rhg
°C
Hel
p
Menu End T
set
T
fix
T il (min) -50.0 °C
T ih (max)302.0 °C
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
rhK
°C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda RP 890 and is the answer not in the manual?

Lauda RP 890 Specifications

General IconGeneral
BrandLauda
ModelRP 890
CategoryThermostat
LanguageEnglish

Related product manuals