Proline Low-temperature Thermostats
YACE0072 / 21.08.07 Starting up 71
7.8.10 Restoring the works setting of the external temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
cDE D
3x to LMDls
2x
to B@K-- 3x
Call the offset works setting for the external temperature probe
cDE D .
Continue as described in Ì 7.8.8 for the internal temperature
probe.
Command
Default
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Æ Settings Æ Calibration Æ
extern Pt100 Æ Default .
The adjacent window appears.
Confirm selection with .
Continue as described in 7.8.8 for the
internal temperature probe.