EasyManua.ls Logo

LEAD FLUID BT600S - 8 Maintenance; Warranty; Regular Maintenance; Malfunction Solutions

LEAD FLUID BT600S
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LEAD FLUID TECHNOLOGY CO.,LTD.
between 0.1 and 100 rpm; the speed resolution is 1 rpm when the speed is
between 100 and 600 rpm. The speed can be set by pressing the UP or
DOWN key. Press the UP or DOWN key shortly, the last digit of the value
will change 1. Hold the UP or DOWN key to change the value fast.
Hold the PRIME key, then press the UP key to set the speed to maximum
directly; hold the PRIME key, then press the DOWN key to set the speed to
minimum (0.1 rpm) directly.
8 Maintenance
8.1 Warranty
The product comes with one-year labor and parts warranty. The limited
warranty does not cover any damage that is caused by improper usage and
handling.
8.2 Regular Maintenance
1) Always check the tubing and connections to make sure there is no
leakage.
2) Do not cover the fan on the rear of the pump.
3) Do not use water to wash the pump. Keep pump head dry.
4) Do not use chemical solvents to clean pump and pump head.
8.3 Malfunction Solutions
No.
Malfunction
Description
Solution
1 Hardware No display 1. Check the power cord
2. Check the fuse. If it was blown,
replace it with a 1A slow-blow fuse
3. Check the internal power cord
connection inside the pump.
2 Hardware Motor does
not work
1. Check the indicator of the driver
board.
Mailmaster@leadfluid.com WEBwww.leadfluid.com 22