ASP6025 Tissue Processor
4. Instrument Setup
4.2 Basic instrument/hardware
Fig. 20
• Itistheuser'sresponsibilitytomakesurethatacompatibleelectromagneticenvironment
is maintained so that the instrument can work as intended.
• Theinstrumentrequiresaninstallationareaofapprox.700x800mm.
• Thesubstratemusthaveasucientloadcapacityandrigiditywithrespecttotheweight
of the instrument.
• Relativehumidity80%maximum-non-condensing.
• Roomtemperatureconsistentlybetween+15°Cand+40°C.
• Elevation:Uptoamax.of2000mabovesealevel.
• Ambientpressurefrom740hPato1100hPa.
• Theinstrumentisdesignedforindooruseonly.
• Thepowersupplymustbewithintherangeofthelengthofthepowercable.Noextension
cable may be connected.
• TheinstrumentMUST be connected to a grounded power socket.
• Useonlyoneoftheprovidedpowercablesthatisintendedforthelocalpowersupply.
• Avoidvibrations,directsunlightandheavyvariationintemperature.
4.2.1 Installation site requirements
CAUTION!
The chemicals to be used in the
ASP6025
are easily inammable and hazardous to health.
Therefore the installation location must be well ventilated, and must contain no open ames. The room in
which the instrument is located must not be used as a permanent station for people. Otherwise the room must
be equipped with an exhaust device.
The installation location must be protected against electrostatic discharge.
The instrument must be set up so that the power switch on the rear side of the instrument (Fig. 3, item 37) and
the power plug are easily accessible at all times.
Do not operate the instrument in rooms with explosion hazard.
To ensure proper functioning of the instrument, it must be set up while maintaining a minimum distance of
10 cm from walls and furniture.
After unpacking the instrument, the instrument must be handled only by
the handles (26) on the rear side (Fig. 3) to move it to its nal location. After-
wards, the brakes must be applied to the instrument's castors.