EasyManua.ls Logo

Leica DMi1 - Technical Description; Objective Specifications

Leica DMi1
132 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
11. Technical Description
11. Technical Description
Due to basic principles of physics and the physi-
ology of the eye, all methods – not just those in
microscopy – have performance limits. There-
fore, observe the following information for cor-
rect use of the microscope.
Performance data of the objectives
The Leica DMi1 microscope is based on the tube
length ∞ (infinity) and a tube lens focal length of
f = 200 mm.
!
Caution:
Therefore, only objectives with the engraving ∞
and the thread dimension M 25 may be used.
Objective marking
Examples and meaning of the symbols:
∞ / –
HI PLAN 10x/0.22
∞ / 0.17
N PLAN 40x/0.65
∞ / 0 / D
N PLAN 50x/0.75
Objective for infinite tube length (∞).
The objective can be used with and
without cover slip.
0.17 The objective may be used with a cover
slip of the standard thickness 0.17 mm
only. If the cover slip is missing or the
thickness of the cover slip deviates
greatly, significantly decreased perfor-
mance will result, particularly for high
objective apertures (see below).
0 Application without cover slip, e.g. for
cell smears, incident light. Not suited
for inverted microscopes.
D or
(A, B, C) Eyepoint of the objective (important for
Modulation Contrast IMC, for example)
L Long working distance.
10x/0.22 Magnification and aperture. The aper-
ture determines the resolution, focus
depth, contrast and brightness of the
microscope image. Objectives with
built-in iris diaphragm have the maxi-
mum and minimum aperture engraved,
e.g. 0.85-0.55.

Table of Contents