EasyManuals Logo

Leica DMi8 A User Manual

Leica DMi8 A
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
2. Safe t y NoteS
Leica DMi8 M microscope (manual)
For indoor use only.
Supply voltage: 100 – 240 VAC
Frequency: 50 / 60 Hz
Power consumption: max. 55 VA
Fuses: 1.6 A, slow-blowing,
breaking capacity H, 250 VAC
Size: 5x20 mm
Ambient temperature: 15° - 35°C
Relative humidity: 90% up to 30°C,
non-condensing
Protection class: I
Overvoltage category: II
Pollution degree: 2
Leica DMi8 C microscope (manually encoded)
For indoor use only.
Supply voltage: 100 – 240 VAC
Frequency: 50 / 60 Hz
Power consumption: max. 55 VA
Fuses: 1.6 A, slow-blowing,
breaking capacity H, 250 VAC
Size: 5x20 mm
Ambient temperature: 15° - 35°C
Relative humidity: 90% up to 30°C,
non-condensing
Protection class: I
Overvoltage category: II
Pollution degree: 2
Leica DMi8 A microscope (automated variants)
For indoor use only.
Supply voltage: 100 – 240 VAC (Leica CTR)
Frequency: 50 / 60 Hz (Leica CTR)
Power consumption: see Leica CTR
Fuses: see Leica CTR
Ambient temperature: 15° - 35°C
Relative humidity: 90% up to 30°C,
non-condensing
Protection class: I (Leica CTR)
Overvoltage category: II (Leica CTR)
Pollution degree: 2 (Leica CTR)
Leica EL6000*
For indoor use only.
Supply voltage: 100 – 240 VAC
Frequency: 50 / 60 Hz
Power consumption: max. 200 VA
Fuses: 2.5 A, slow-blowing,
breaking capacity H, 250 VAC
Size: 5x20 mm
Ambient temperature: 0° - 40°C
Relative humidity: 90% up to 30°C,
non-condensing
Protection class: I
Overvoltage category: II
Pollution degree: 2
(* See provided Instructions for Use)
7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Leica DMi8 A and is the answer not in the manual?

Leica DMi8 A Specifications

General IconGeneral
BrandLeica
ModelDMi8 A
CategoryMicroscope
LanguageEnglish

Related product manuals