3.6 Calibrating the Encoders
☞
Calibration of the wheel encoder when using the DSX for the rst
time is mandatory.
We recommend, to calibrate the wheel encoders at least once on
each scan site. In particular when ground conditions change - for
example, from asphalt to grass.
1. Measure out a distance of exact 10 m. The dened distance is used
as a reference for the calibration procedure.
☞
Ensure that the tablet is connected to the DSX.
2. Turn on the DSX. Turn on the tablet and start the DXplore software.
3. Start the calibration procedure with the DXplore software. Refer to
Wheel encoder calibration.
4. Move the DSX along the dened distance.
5. Complete the calibration procedure with the DXplore software and
check the calibration result. If necessary, repeat the calibration.
Calibration procedure
24 Setup