Appendix 1
Envisu C-Series SDOIS / 9054-10011 / Revision N 29
The Models Envisu 2300 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Envisu 2300 Model should assure that it is used in such an environment.
Electromagnetic environment
guidance
2300 Model requires continued
operation during power mains
interruptions, it is recommended that
the Envisu 2300 Model be powered
from an uninterruptible power supply or
a battery.
Rated Power frequency
magnetic field
IEC 61000-4-8
If image distortion occurs, it may be
necessary to position the Envisu 2300
Model further from sources of power
frequency magnetic fields or to install
magnetic shielding. The power
frequency magnetic field should be
measured in the intended installation
location to assure that it is sufficiently
low.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.