Check & Adjust TPS1200+ 91
Every angle measured in the daily work is corrected automatically if the compensator
and the Hz-corrections are activated in the instrument configuration. Select Main
Menu: Config...\Instrument Settings...\Compensator to check the settings.
View current
adjustment errors
The currently used adjustment errors can be viewed under Main Menu:
Tools.../Check & Adjust...\Current Values.
Mechanical
adjustment
The following instrument parts can be adjusted mechanically:
Precise
measurements
To get precise measurements in the daily work, it is important:
To check and adjust the instrument from time to time.
To take high precision measurements during the check and adjust procedures.
To measure targets in two faces. Some of the instrument errors are eliminated
by averaging the angles from both faces.
Refer to "4.2 Preparation" to find more important points.
Circular level on instrument and tribrach
Laser plummet
Optical plummet - option on tribrach
Allen screws on tripod

Table of Contents