92TPS1200+
Check & Adjust
During the manufacturing process, the instrument errors are carefully determined
and set to zero. As mentioned above, these errors can change and it is highly recom-
mended to redetermine them in the following situations:
Before the first use
Before every high precision survey
After rough or long transportations
After long working periods
After long storage periods
If the temperature difference between current environment and the temperature
at the last calibration is more than 20°C