To
Load
In
order
to
help
avoid
inadvertent
infusion
of
air
to
the
patient,
the
following
warnings
should
be
observed:
|
WARNING
!:
Remove
all
air
from
solution
bags.
|
WARNING
|:
Not
for
use
with
commercial
I.V.
solution
bags
less
than
1
liter
with
air.
|
WARNING
|:
Not
for
use
with
autotransfusion
bags.
1.
Set
up
and
prime
Level
1
D-series
disposable
|.V.
fluid
warming
set
following
the
instructions
on
disposable
package.
2.
Flip
toggle
switch
on
the
top
front
of
the
H-2
chamber
all
the
way
over
to
the
“-
OFF
position.
3.
Open
chamber
door
by
swinging
out
hinged
latch
on
right
side.
4.
Hang
air-free
solution
bag
on
door
post.
NOTE:
Each
pressure
chamber
can
hold
either
a
500
ml
bag
or
a
1000
ml
bag.
Fora
500
ml
bag,
the
hanging
post
is
located
on
the
inside
of
the
door.
For
a
1000
ml
bag
the
post
is
located
on
the
top
of
the
door.
For
optimum
drainage
of
1liter
bags,
the
upper
molded
post
is
for
longer
style
bags.
The
lower
molded
post
is
used
for
standard
size
bags.
5.
Close
door
and
secure
side
latch.
To
Pressurize
CAUTION:
Be
certain
latch
is
secure
before
chamber
is
pressurized.
Never
pressurize
with
door
open.
CAUTION:
Do
not
exceed
300
mmHg.
1.
Push
toggle
switch
all
the
way
over
to
the
“+”
or
ON
position.
Observe
pressure
gauge
to
ensure
300
mmHg
(approx.)
is
achieved.
NOTE:
The
pressure
is
not
adjustable
on
this
device.
The
fluid
warmer
power
switch
must
be
ON
for
pressure
chambers
to
operate.
To
Unload
| |
1.
Push
toggle
switch
all
the
way
over
to
the
“-“
or
OFF
position.
This
will
release
pressure
in
the
chamber
and
deflate
the
bladder.
2.
Open
door
and
remove
the
fluid
bag.
11