10. MJB8 (MEDLINE
®
CARELINE
®
)
The MJB8 modular junction box is a central unit in the “Intelligent Care Bed”, connecting various
intelligent accessories, such as the Out Of Bed and WET detection, and sending notifications via a range
of optional gateways to the user (i.e. caregiver)
Recommendations
Always use locking mechanism and O-ring.
Sockets that are not used must be fitted with blind plugs (Item P/N.: 0821008) to ensure IP-degree.
When using the modular plug cable with an open end, the customer is responsible for maintaining the IP degree.
When mounting, ensure that a screw torque limit of 1 Nm is not exceeded.
• 3
rd
party products must correspond to all requirements stated in the MJB8 Interface description for 3
rd
party products in order to avoid damage/
malfunction.
The Under Bed Light (Item P/N.: 0964135) must be mounted on the bed with metal screws in order to maintain ESD protection.
HOT PLUGGING
Removing or adding any OpenBus™ cables are not allowed when the control box is powered by mains supply!
Follow the below procedure:
1. Remove mains and wait 5 seconds
2. Mount or dismount the required cables
If this procedure is NOT followed it may result in a damaged OpenBus
TM
driver circuit.
The risk of a damaged circuit increases if the accessory has a high start current.
Before the final functional production test, it is important that the system is repowered.
This is to ensure, that all items have been detected on the OpenBus
TM
.
It is important to test the specified notifications in order to ensure that they work correctly before sending the system to the end user.
The MJB8 is intended for mains operation only. The user must be informed that all MJB8 sensor notifications are disabled when the system is in
battery mode.
Warnings
LINAK only takes responsibility for LINAK products, not 3
rd
party products.
Please pay attention to the “Patient Environment” Clause 3.79 – EN60601-1 3
rd
edition. It must be subject to the Risk Analysis.
It is important to inform the customer about this.
The MJB8 is not able to detect defective 3
rd
party products.
We recommend the end user to make a regular test procedure in order to prevent hazardous situations for the user and failures to the system.
The MJB8 and the attached accessories (i.e. Out Of Bed or WET detection) are not intended as life-supporting or emergency equipment.
They are only intended to support notifications for communication and comfort purposes of people in need of care.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Linak CB14 and is the answer not in the manual?

Linak CB14 Specifications

General IconGeneral
BrandLinak
ModelCB14
CategoryController
LanguageEnglish

Related product manuals