Connection of Emergency lowering:
Please observe that if you order the TD3 with 2 x 9V batteries mounted, the
batteries will not be connected. The customer needs to connect the cable from
the battery drawer to the TD3 when installing the bed and connecting the
mains.
If the batteries are connected and the bed is not connected to the mains,
the batteries will discharge in approx. 2 months and it will not be possible to
perform an emergency lowering.
Lifetime of batteries:
The batteries must be changed after an emergency lowering has been
performed. Batteries that have not been used must be changed after approx.
3 years.
Battery type:
Alkaline 9 V batteries 9F22 or 6LR61.
The correct way to use the functionality:
If the key is released during the lowering, always wait 60 sec before pressing the key
again.
Removal/fi tting of the 2 x 9 V batteries:
To change the 2 x 9 V batteries for emergency lowering the fi rst step is to disconnect
the cable and then pull out the drawer of the TD3. On the bottom of the drawer,
there are 2 holes, which are to be pushed with a fi nger or a blunt instrument. The
batteries will be released from the drawer when pushing. The new batteries can be
placed in the drawer. Afterwards, place the drawer in TD3 and connect the cable
again.
Only use alkaline batteries, and make sure to polarise the batteries correct according
to the symbols in the drawer.
To pull out the batteries
press here
Remember to
connect the cable
again afterwards